Model Number NT4W18115 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/12/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus and will be sent to the oem for evaluation.The cause of the reported event cannot be determined; however, the mostly likely cause of the reported event can be attributed to a higher than normal force applied on the joint by the operator when extracting a large stone.The instruction manual warns users: certain objects may be too large to remove endoscopically with this retrieval system.To avoid complications fluoroscopic x-ray or some means of identification of an objects size should be used prior to attempting to remove.Do not use this device if the object is too large to be removed and/or cannot be securely held in the stone dislodger.Inspect the device for any visible damage such as kinks or broken stone dislodger wires.Test the device by opening, closing and rotating it several times before being introduced into the patient.
|
|
Event Description
|
Olympus was informed that during a laser lithotripsy procedure, the basket broke off and fell into the patient.The physician used another device and retrieved the basket.It was reported that the physician¿s retrieval delayed the procedure 45 minutes.There was no bleeding reported.The intended procedure was completed with a different device.There was no patient injury.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the device evaluation results.The device was returned to the oem for investigation.A visual inspection of the device was performed and found that the basket was not retuned; however, the remainder of the device was coiled up with the thumb slide in a partial deployed position.The hypo tube was still attached to the core wire and about 0.234¿ from the end of the sheath.During visual inspection it was noted that there were 2 kinks in the sheath; one at 2.75¿ and the other 0.141¿ from the distal end of the sheath.The welds on the hypo tube were inspected under a microscope and no anomalies were found as the welds measured within manufacture¿s specification.A review of the device history reveals that there were no anomalies noted during the manufacturing of this device.Due to the basket not being returned and based on the investigation findings, the most likely cause of the reported event is either the welds on the hypo tube did not penetrate deep enough to secure the basket wire bundle or the basket wire bundle did not get located properly in the hypo tube prior to the welding process.
|
|
Search Alerts/Recalls
|