• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Charging Problem (2892); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that a medical professional was having issues with his handheld computer.It was reported there was an issue with the touch screen and to charge the device.Review of manufacturing records confirmed all tests passed for the device prior to distribution.
 
Event Description
The suspected handheld computer and the flashcard were returned to the manufacturer on 10/13/2016.An analysis of the returned flashcard was completed and no anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.An analysis of the returned handheld computer was completed and the reported allegation of computer screen responsive issue¿ was not verified.No anomalies associated with the handheld touchscreen were identified during the analysis.An analysis was performed on the returned handheld and the reported allegation of ¿battery failure¿ was verified.An analysis of the handheld identified that the returned battery was swollen, and unable to power the handheld for an hour.Once the battery was replaced with a known good battery, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5999207
MDR Text Key56447218
Report Number1644487-2016-02160
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number840997
Other Device ID NumberVERSION 8.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 09/13/2016
Initial Date FDA Received10/05/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/31/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-