The suspected handheld computer and the flashcard were returned to the manufacturer on 10/13/2016.An analysis of the returned flashcard was completed and no anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.An analysis of the returned handheld computer was completed and the reported allegation of computer screen responsive issue¿ was not verified.No anomalies associated with the handheld touchscreen were identified during the analysis.An analysis was performed on the returned handheld and the reported allegation of ¿battery failure¿ was verified.An analysis of the handheld identified that the returned battery was swollen, and unable to power the handheld for an hour.Once the battery was replaced with a known good battery, no further anomalies were identified.
|