(b)(4).The avalon cannula was returned to the factory and investigated in the decontamination / complaints laboratory.The cannula was delivery in a damaged condition to the laboratory.The cannula was found to be almost completely broken through, with only the wire and a small piece of tubing holding the cannula together.The customer's reported issue could be confirmed.The issue was referred to the responsible r&d engineer, who also discussed the complaint with the manufacturer, avalon.According to the evaluation by r&d and avalon, the cannula has been cut on the proximal side during the process of fixation with a suture during normal use.The instructions for use (ifu) g-158 (art.# 70106.3902, common device name: ifu: avalon elite, dual lumen cath.G-158, version 02) clearly states the following: item 5: instructions for use, subitem 5 "securing the device: the catheter should be secured to the patient around the rigid barbed connectors in manner that does not kink the catheter at the insertion site.Caution! a suture tied directly around the wire-reinforced section can cut, kink, or damage the catheter." the conclusion from the investigation that damage to the catheter was as a result of use not in accordance with the ifu.This is a user-caused, and not a product-related, issue.Based on the investigation and trending for this issue, a systemic issue is not indicated, therefore no further investigation or action is currently warranted for this issue in addition to continued periodic monitoring.
|