The pipeline flex pushwire and the microcatheter were returned.The pipeline flex pushwire was found outside of the microcatheter.The pipeline braid was not returned as it was implanted in the patient.It was confirmed that the pipeline flex pushwire was detached at the hypotube proximal to the wire weld.The distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.The pushwire was also observed to be bent.The microcatheter was found to have accordion damage.No other anomalies were observed.The surfaces of the detached push wire were sent for sem (scanning electron microscope) and eds (energy dispersive spectroscopy) analyses.Based on the analysis findings, the clinical observation was confirmed.It appears that the distal wire was secondary to tensile failure.We are unable to definitively determine for the tensile failure.The damages seen on the returned proximal wire, hypotube and catheter body suggest excessive forced was used such as pushing and pulling.The elemental analysis conducted through scanning electron micrographic (sem) / energy dispersive spectroscopy (eds) showed presence of soldering material; thereby indicating that the soldering was conducted.In addition, a review of this device lot history record was conducted and no issues were identified that could have contributed to this event.All products are 100% inspected for damage and irregularities during manufacture.Per our instructions for use, the user should "discontinue delivery of the device if high force or excessive friction is encountered.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.
|
Medtronic received information that during treatment of an aneurysm, located in the left cavernous segment of the internal carotid artery the push wire tip coil detached distal of the hypotube with the ptfe sleeves still attached.It was reported that the pipeline was implanted successfully; however when the microcatheter was used to withdraw the push wire tip coil, resistance was felt.The pipeline and microcatheter were then withdrawn together from the patient and the tip coil was discovered to be separated.The tip coil was still within the microcatheter and removed from the patient successfully.There was no report of patient injury.
|