MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARDEX 2-WAY 5CC ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARDEX 2-WAY 5CC ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 165814
Device Problems Break (1069); Entrapment of Device (1212); Difficult to Remove (1528)
Patient Problems Pain (1994); Discomfort (2330)
Event Date 08/17/2016
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the catheter formed a ridge when removing it from the patient; as a result, the ridge made it very difficult to remove the catheter due to it being stuck.The complainant reported that the catheter was pulled using force and caused the tip to break off inside the patient.The tip of the catheter was manipulated out of the patient without any additional interventions.The removal of the catheter allegedly caused the patient pain and discomfort.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX 2-WAY 5CC ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
Manufacturer (Section G)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX   85621
Manufacturer Contact
janna parks
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6004476
MDR Text Key57255197
Report Number1018233-2016-01360
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/27/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number165814
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/27/2016
Initial Date FDA Received10/06/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-