It was reported that the external pulse generator stopped working and presented with an error code.After the external pulse generator was re-booted, it then presented with a different error code.The external pulse generator has been returned for repair, and a loaner was requested.There was no patient involvement.
|
Failure analysis was performed on the main printed circuit board assembly.Visual inspection revealed no anomalies.Bench analysis found that the device powered up normally.The device was run on the automated test console, all tests passed.The device was placed in an oven at 158f overnight, no error then put in the freezer for an hour, no error.The device was then put directly back in the oven, when checked an hour later an error was present.Additional analysis: the error log was reviewed, the log showed multiple errors.The errors were for a timeout waiting for event latch write by die stack.Conclusion: confirmed the reported event, suspected solder issue with the die stack.
|