• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems Failure to Pump (1502); Device Displays Incorrect Message (2591); Human Factors Issue (2948)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/09/2016
Event Type  No Answer Provided  
Manufacturer Narrative
Patient information was not provided.(b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the s5 roller pump displayed an error code during a procedure.There was no report of patient injury.A sorin group field service representative contacted the customer to provide tech support.The customer stated that, after receiving the error message, the pump was switched out and the procedure was continued with no additional problems.During the technical support call, the service representative learned that the issue was caused by an operator error.The involved pump was set as a cardioplegia pump utilizing the dual tubing method in the raceway.The tubing got twisted and resulted in the reported error message.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.No trend was identified for this type of issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.Technical support provided via phone.
 
Event Description
Sorin group (b)(4) received a report that the s5 roller pump displayed an error code during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
carrie wood
14401 w. 65th way
arvada, CO 80004
3034676461
MDR Report Key6012762
MDR Text Key56964183
Report Number9611109-2016-00639
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 09/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/09/2016
Initial Date FDA Received10/07/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-