• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 1/8" PEG DRILL; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 1/8" PEG DRILL; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 6541-4-518
Device Problems Break (1069); Fracture (1260); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 09/09/2016
Event Type  malfunction  
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Event Description
Triathlon instruments broke during surgery.
 
Manufacturer Narrative
An event regarding crack/fracture involving a 1/8¿ peg drill was reported.The event was confirmed.Method & results: -device evaluation and results: review with the material analysis engineer indicated "examination of the peg drill indicated it fractured due to a torsional overload condition" part of the peg drill tip has been fractured off.-medical records received and evaluation: not performed as medical records were not provided.-device history review: review of the device history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.-complaint history review: there have been 2 other events for the lot referenced.Conclusions: the investigation concluded that the 1/8¿ peg drill fractured due to a torsional overload condition which was determined by the mar team.
 
Event Description
Triathlon instruments broke during surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1/8" PEG DRILL
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6013540
MDR Text Key57361078
Report Number0002249697-2016-03190
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K143393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6541-4-518
Device Lot NumberSC3A01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/13/2016
Initial Date FDA Received10/10/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/17/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/06/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-