• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PULSE GEN MODEL 106; GENERATOR Back to Search Results
Model Number 106
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Staphylococcus Aureus (2058)
Event Date 08/18/2016
Event Type  Injury  
Event Description
It was reported on (b)(6) 2016 that the patient had a recent generator replacement on (b)(6) 2016 and developed a staph infection at his generator site afterwards.It was stated that the patient had a full explant (b)(6) 2016 and the devices were disposed after explant.
 
Manufacturer Narrative
Suspect device udi: (b)(4).
 
Event Description
Notes were received on 10/31/2016 and dated (b)(6) 2016.Notes state that he had his generator replacement two weeks ago and since that time has been having frequent seizures for no apparent reason.The vns was checked that day and they saw he was programmed to the same settings as it was at prior to having a new battery implanted.It was noted that he is on rapid cycling so autostim cannot be programmed on at this time due to the setting constraints.No additional relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6015252
MDR Text Key56926019
Report Number1644487-2016-02316
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/15/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/15/2018
Device Model Number106
Device Lot Number203696
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 09/15/2016
Initial Date FDA Received10/10/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age29 YR
-
-