• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY AXIOM ARTIS ZEEGO; INTERVENTIONAL, FLUOROCOPIC, XRAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY AXIOM ARTIS ZEEGO; INTERVENTIONAL, FLUOROCOPIC, XRAY SYSTEM Back to Search Results
Model Number 10280959
Device Problems Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/09/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zeego system.After transferring a patient to the table in preparation for a tavr procedure, a "no communication" error was received and there was no stand movement.The patient was safely removed from the system and we are unaware of any impact to the state of health of the patient involved.
 
Manufacturer Narrative
During the investigation it was found that an ethernet switch caused the system error.After replacement of the ethernet connection, the system works as specified.The root cause for defective switch could not be determined, however, the connection was reestablished the failure did not reoccur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOM ARTIS ZEEGO
Type of Device
INTERVENTIONAL, FLUOROCOPIC, XRAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd
65-1a
malvern, PA 19355
6104486461
MDR Report Key6016622
MDR Text Key56980967
Report Number2240869-2016-50930
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K141574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 09/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10280959
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/09/2016
Initial Date FDA Received10/11/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-