Model Number 103
Device Problems High impedance (1291); Positioning Problem (3009)
Patient Problems Dyspnea (1816); Tingling (2171)
Event Date 09/15/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the patient had gone to the er on (b)(6) 2016 complaining of dyspnea, coughing, and a tingling sensation in her face and upper extremities.The physician tested the vns and found high impedance.The vns was programmed off and the patient was referred for surgery.The patient underwent surgery to correct the high impedance observed, on (b)(6) 2016.It was noted that the lead pin was not fully inserted into the generator.The lead pin was reinserted; the impedance was checked both outside and inside the generator pocket.Both sets of diagnostics were within normal limits and there was no need for a replacement of either the lead of the generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6019996
MDR Text Key57111010
Report Number1644487-2016-02335
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/14/2017
Device Model Number103
Device Lot Number4351
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 09/16/2016
Initial Date FDA Received10/12/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/21/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-