Evaluation summary: post market vigilance (pmv) led an evaluation of one device opened by the account.This evaluation was based on a technical review of all data received from the site, a pmv review of manufacturing records, a pmv review of complaint trends, and an evaluation of the returned device.The reported condition for this incident was that during bariatric procedure there was a device power failure.The instrument was connected to pmv equipment and registered a mechanical fault.Visual inspection of the instrument noted the probe tip was broken.Visual and functional testing of the returned product confirmed the product did not meet quality release specifications that were tested regarding the reported conditions due to the broken tip.The file was concluded to be a misuse of the product.Replication of the observed damage may occur when allowing the energized probe tip to come in contact with other metal objects such as other instruments, staples, or clips.Doing so may generate excess heat and weaken or break the probe tip.The instructions for use of this instrument state: avoid all contact between the tip of the instrument and metallic objects as it may cause damage and render the instrument inoperable.The event did not meet the regulatory reporting criteria.A review of the current historical complaint data reveals no trend for a device related failure for this condition.The file will be closed as a misuse of the product.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.
|