Catalog Number 254500506 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Information (3190)
|
Event Date 10/07/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
The trial is missing one of the balseals.
|
|
Manufacturer Narrative
|
The device associated with this report was not evaluated.Hhe/qrb (quality review board) (b)(4) recommended a device correction which was initiated on june 12, 2015.Capa-(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.Capa-(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor per post market surveillance sep-(b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
The investigation has been reopened due to receiving product for evaluation.Depuy will notify the fda when the investigation is complete.
|
|
Manufacturer Narrative
|
Examination of the returned device confirms the reported event of damaged/missing balseals.(b)(4) recommended a device correction which was initiated on june 12, 2015.(b)(4) determined the likely root cause to be related to misuse, and requires mandatory sales training on proper use and application of the device by depuy sales consultants.(b)(4) will also monitor the mandatory field training.The need for further corrective action was not indicated.Continue to monitor disassociation events per post market surveillance (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|