• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ACL TIGHTROPE RT; PIN, FIXATION, SMOOTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ACL TIGHTROPE RT; PIN, FIXATION, SMOOTH Back to Search Results
Catalog Number AR-1588RT
Device Problem Patient-Device Incompatibility (2682)
Patient Problems Hypersensitivity/Allergic reaction (1907); Rash (2033); Swelling (2091); Burning Sensation (2146); Reaction (2414)
Event Date 08/04/2016
Event Type  Injury  
Manufacturer Narrative
Patient demographics (age at time of event, date of birth, weight) were requested but not provided.This is the fifth of four submissions from the same patient event.The others are 1220246-2016-00359 ((b)(4)), 1220246-2016-00360 ((b)(4)),1220246-00361 ((b)(4)) and 1220246-2016-00362 ((b)(4)).No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.Lot number was not provided so device history record review cannot be performed.The device was not returned for evaluation but remained in the patient therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.The most likely cause of this type of event is a reaction of the patient to the material(s) implanted.Product directions for use warns of effects like foreign body and allergic-like reactions as well as infections both deep and superficial to the implant materials.Patient sensitivity to device materials must be considered prior to implantation.Device remains in patient.
 
Event Description
It was reported by the patient on day of the revision procedure, (b)(6) 2016, that she had undergone an acl reconstruction on (b)(6) 2016 on her right leg.On (b)(6) 2016 the patient began to notice extreme swelling, knee was hot to the touch, she had a skin rash with welts and burning in and around the incision.The reaction also spread to the lateral side of the femur and tibia.The patient had an er visit on (b)(6) 2016 where bloodwork was taken.Results were negative for infection.Surgeon decided to bring patient in for second surgery and possible removal of the implants.Patient states that prior to the (b)(6) 2016, procedure she had allergy testing which showed that she was allergic to nickel.The following parts were implanted during the original (b)(6) 2016 procedure: ar-1934bcf lot 1240957 ((b)(4)), ar-1934bcf lot 1240946 ((b)(4)), ar-2324bcc lot 10025466 ((b)(4)) and ar-5035tc-09 lot 1314851 ((b)(4)).Additional information was obtained on (b)(6) 2016 as follows: the patient's mother called to report that during a surgeon visit while they were going over the material composition of the patients implants listed on the facility implant log the surgeon informed them that there was an additional part that was used for the patient's surgery that he noticed the or staff did not include on the patient's implant log.The missing part number provided to the patient by the surgeon was ar-1588rt, acl tightrope rt with titanium and uhmwpe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE RT
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath, sr mdr analyst
1370 creekside boulevard
naples, FL 34108-1945
8009337001
MDR Report Key6030707
MDR Text Key57553406
Report Number1220246-2016-00428
Device Sequence Number1
Product Code HTY
UDI-Device Identifier00888867019553
UDI-Public00888867019553
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Patient
Type of Report Initial
Report Date 09/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberAR-1588RT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/20/2016
Initial Date FDA Received10/14/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-