X-rays were received to review if the patient had a lead fracture that might have been contributing to the reported events.The review found that the lead pin was properly inserted, adequate strain relief was present, and no gross fractures could be visualized.However the possibility of micro fractures that could not be visualized with naked eye were not ruled out.The patient's lead and generator were then explanted on (b)(6) 2016 and during the explant surgery it was discovered the patient had an infection despite having a negative cbc count.The infection appeared to be wide spread extending from the generator site to the lead site and causing the drainage at the electrode site.During surgery there was no nerve damage noted which the surgeon had mentioned as a possibility prior.The patient had antibiotics prescribed to treat the infection and had the entire vns system removed.The vns system was then discarded by the explanting facility.The surgeon and physician did not know what caused the infection.Patient behavior was discussed as a possibility but no conclusive decision was found.The device history record for the generator and lead were reviewed and both devices were found to be sterilized prior to shipment.The expiration dates were 01/18/2019 and 01/22/2017 for the lead and generator, respectively.No other relevant information has been received to date.
|