Model Number 106
Device Problems Break (1069); Improper or Incorrect Procedure or Method (2017)
Patient Problems Fall (1848); Unspecified Infection (1930); Organ Dehiscence (2502)
Event Date 09/29/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the patient had her vns explanted on (b)(6) 2016 as her incision had broken open.The surgeon was concerned that everything was infected.Originally, it was unknown how the dehiscence occurred; however, it was later found that the patient had fallen on an unknown date, which caused the generator site incision to open.Review of the dhrs for both the lead and the generator confirmed sterilization prior to distribution.
 
Event Description
It was later reported that after the patient's device was explanted she began to have an increase in seizures.The increase in seizures was caused by the loss of therapy as the device was explanted due to the infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6036101
MDR Text Key57686269
Report Number1644487-2016-02390
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/19/2018
Device Model Number106
Device Lot Number203837
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 09/30/2016
Initial Date FDA Received10/18/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/21/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
-
-