The hvad is used for treatment not diagnosis.The hvad pump driveline cable ((b)(4)) was not returned for analysis because it remains implanted.The reported event, driveline wire damage, was confirmed visually in the field by a heartware clinical engineer.It is possible that when the patient dropped the controller, the tensile force may have caused the pins to be disconnected from the contact block.After the driveline splice procedure was performed, the power and flow returned to normal and the patient is reported to be doing fine.The root cause for the reported driveline connector damage can be attributed to inadequate connector design specifications that does not account excessive pull force that may be exerted on to the driveline cable, causing the connector pins to recess and interrupt pump operation.The manufacturer has an open internal investigation for the retraction of the driveline cable splice connector pins.The ventricular assist system is indicated for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings, including transportation.Per ifu provides an instruction and also a caution note: do not pull, twist or kink the driveline or power cables, especially while sitting, getting out of bed, adjusting controller or power sources, or when using the shower bag.It also cautions, "keep extra driveline length tucked under clothing or secured with an abdominal binder or dressing.Do not let any portion of driveline hang freely where it might get caught on external items such as doorknobs or the corners of furniture.Heartware is submitting this correction to a previously submitted mdr report as a result of remediation activities related to fda warning letter fla-14-14, dated june 2, 2014, and pursuant to the provisions of 21 cfr part 803.(b)(4).
|