Model Number FUSION COMPACT
Device Problems Device Stops Intermittently (1599); Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/22/2016
Event Type  malfunction  
Manufacturer Narrative
Return requested.Replacement computer shipped to site 09/29/2016.No parts have been received by manufacturer for analysis.No further issues have been reported.
 
Event Description
A medtronic ent representative reported that, while in an ear, nose & throat (ent) procedure, a balloon sinus dilation, the axiem box was not working with the navigation system.When in the software, the emitter details showed both the emitter and axiem box as red and not communicating.In trouble-shooting, re-booted in multiple states of having the axiem box plugged in before/after re-boot and in different ports, however, that and other steps did not resolve the issue.The surgeon opted to abandon the use of the navigation system to continue the procedure to completion.Delay in therapy was 5 minutes.There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key6042193
MDR Text Key57923913
Report Number1723170-2016-02633
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 10/19/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/22/2016
Initial Date FDA Received10/19/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age52 YR
Patient Weight84
-
-