Catalog Number 865226 |
Device Problem
Break (1069)
|
Patient Problem
No Information (3190)
|
Event Date 09/27/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Extractor shaft has come away from handle.Threaded tip has snapped off.
|
|
Manufacturer Narrative
|
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Manufacturer Narrative
|
Conclusion and justification status for mdr: examination of the submitted 865226 pfc stem trial extract confirmed the threaded tip has broken off and was not returned.The investigation also confirms the breakage of the shaft from the handle.The observed breakage is consistent with ductile overload.The root cause of the ductile overload is attributed to suspected levering of the instrument side to side while attempting to remove the fluted rod from the patient bone canal.No evidence was found indicating product error was a contributing factor and the need for corrective action was not indicated.Continue to monitor via sep-419.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|