• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE HEMODYNAMICS; PROGRAMMABLE DIAGNOSTIC COMPUTER Back to Search Results
Model Number MERGE HEMODYNAMICS 9.40
Device Problem Connection Problem (2900)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Event Date 09/20/2016
Event Type  malfunction  
Manufacturer Narrative
Merge technical support attempted to remotely access the client pc in question, however a message was received that the "server returns destination host unreachable." merge technical support advised the customer to contact their onsite it department for further troubleshooting.The site reported back to merge healthcare that the client pc regained connection to the network and its replication to the server was current.The corrective measures performed by the site's it staff were not disclosed to merge healthcare.However, the customer explicitly stated that the problem occurred at the site and not with merge hemo.
 
Event Description
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2016, a customer reported to merge healthcare that the hemo client pc would not connect to the site's server.Information obtained from the customer revealed that the reported problem resulted in a procedure being cancelled and rescheduled for the next day.Other cases were sent to one of the additional labs at the site.With merge hemo not presenting physiological data during treatment, there is a potential for incorrect treatment that could result in harm to the patient.However, there is no indication from the customer that this issue resulted in harm to patients.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE HEMODYNAMICS
Type of Device
PROGRAMMABLE DIAGNOSTIC COMPUTER
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
michael diedrick
900 walnut ridge drive
hartland, WI 53029
2629123570
MDR Report Key6046200
MDR Text Key58020101
Report Number2183926-2016-00772
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMERGE HEMODYNAMICS 9.40
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/20/2016
Initial Date FDA Received10/20/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-