Model Number MMT-723NAS
Device Problems Device Displays Incorrect Message (2591); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/18/2016
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.
 
Event Description
The customer reported via telephone call that the insulin pump alarmed for a motor error.The drive support cap was sticking out.The customer did not recall the blood glucose reading.She had not pressed on the cap while connected to the insulin pump.She was advised to discontinue use of the insulin pump and to revert to a backup plan per a health care professional's instructions.She was advised that the insulin pump would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
The insulin pump was received unable to prime during prime test and alarmed motor error during basic occlusion test due to loose protruded drive support disk.The motor passed motor test.The insulin pump had minor scratched lcd window, cracked case near the display window corners, broken reservoir tube lip and missing the end cap sticker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6047381
MDR Text Key58465733
Report Number2032227-2016-33724
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/18/2016
Initial Date FDA Received10/21/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age61 YR
-
-