Model Number MMT-723NAS |
Device Problems
Device Displays Incorrect Message (2591); Device Dislodged or Dislocated (2923)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/18/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.
|
|
Event Description
|
The customer reported via telephone call that the insulin pump alarmed for a motor error.The drive support cap was sticking out.The customer did not recall the blood glucose reading.She had not pressed on the cap while connected to the insulin pump.She was advised to discontinue use of the insulin pump and to revert to a backup plan per a health care professional's instructions.She was advised that the insulin pump would be replaced and agreed to return the product for analysis.
|
|
Manufacturer Narrative
|
The insulin pump was received unable to prime during prime test and alarmed motor error during basic occlusion test due to loose protruded drive support disk.The motor passed motor test.The insulin pump had minor scratched lcd window, cracked case near the display window corners, broken reservoir tube lip and missing the end cap sticker.
|
|