Model Number MODEL 250
Device Problems Difficult to Interrogate (1331); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2016
Event Type  malfunction  
Event Description
It was reported that the medical professional was having difficulties to interrogate generators when using his handheld computer.It was reported that the device which is very old, does not communicate anymore with generators.Review of manufacturing records confirmed that the programming computer passed all functional tests prior to distribution.The suspected programming computer was returned to the manufacturer on 10/13/2016.Analysis is underway but it has not been completed to date.
 
Event Description
Analysis of the returned flashcard was completed and no anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.An analysis of the returned handheld computer was completed and the reported allegation was not verified.No anomalies associated with the handheld were noted during testing using the ac adapter or the main battery with a full charge.The handheld performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6050205
MDR Text Key58526525
Report Number1644487-2016-02445
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number874811
Other Device ID NumberVERSION 8.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/05/2016
Initial Date FDA Received10/24/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/31/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-