• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS - HOUSTON LEAD MODEL 302

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS - HOUSTON LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problem High impedance (1291)
Patient Problem Electric Shock (2554)
Event Date 08/29/2016
Event Type  malfunction  
Event Description
It was reported that a patient was having revision surgery.High impedance was identified in pre-op.The patient saw the neurologist on (b)(6) 2016, and the patient reported that she occasionally felt shocks in her neck.The physician ran diagnostics, and high impedance was present.There was no trauma or patient manipulation noted at the patient's appointment.The generator was replaced on (b)(6) 2016, but the high impedance was still present.The lead was then replaced, and the high impedance resolved.The explanted generator and lead were discarded.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6051576
MDR Text Key58612214
Report Number1644487-2016-02452
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2008
Device Model Number302-20
Device Lot Number1300
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 09/30/2016
Initial Date FDA Received10/24/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age28 YR
-
-