This follow-up report is being submitted to relay additional information.Confirmed that this was not an adverse event.No product was returned; visual and dimensional evaluations could not be performed and the complaint could not be confirmed.Both the device history records and the receiving inspection report were reviewed and no deviations or anomalies were identified.It was reported that the surgical technique was followed.However, the exact conditions under which the fracture of the pin occurred remain unknown.This device was produced on 05-feb-2012 and therefore, had been in service for over 4 years.It is unknown how many times the device was used during that time.Per the package insert of this instrument, end of life is determined by wear and damage due to use.Also, the insert recommends to inspect all instruments carefully prior to each use, and to functional test instruments with long slender features (particularly rotating instruments) for distortion.If damage or wear is noted that may compromise the function of the instrument, it should not be used.A definitive root cause cannot be determined with the information provided.
|