• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TAEWOONG MEDICAL CO., LTD. NITI-S PYLORIC & DUODENAL UNCOVERED STENT; PYLORIC & DUODENAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TAEWOONG MEDICAL CO., LTD. NITI-S PYLORIC & DUODENAL UNCOVERED STENT; PYLORIC & DUODENAL STENT Back to Search Results
Model Number DDT2208
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  Injury  
Manufacturer Narrative
Since the device was not returned, it is impossible to proceed to actual evaluation.But, it is confirmed from the device history record that the suspected device had been manufactured with no significant issue and passed all the inspections.It is hard to find out exact root cause for this complaint, because the suspected device was not returned and it is difficult to reconstruct the situation at that time in the lab and limited info.The suspected device is not registered in the us and we will continuously monitor whether similar or same complaint occurs.
 
Event Description
On (b)(6) 2016: ddt2208 was applied to a patient with duodenal cancer.Stenosis was found at the second portion, which was about 2-3 cm in length.During placement, ddt2208 was deployed slightly off the stenosis position so another stent placement (ddt2210) followed.Unknown date: ddt2208 migration was confirmed.On (b)(6) 2016: migrated stent (ddt2208) was found at ileocecal area.Emergency operation took place.Migrated stent was removed during resection of the small intestine.No returned device available for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S PYLORIC & DUODENAL UNCOVERED STENT
Type of Device
PYLORIC & DUODENAL STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO., LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO., LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
sanghyung park
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
19960641
MDR Report Key6053241
MDR Text Key58242580
Report Number3003902943-2016-00033
Device Sequence Number1
Product Code MUM
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 10/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/21/2019
Device Model NumberDDT2208
Device Catalogue NumberTW-TC-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/07/2016
Initial Date FDA Received10/25/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-