BOSTON SCIENTIFIC - MARLBOROUGH SOLYX SIS SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, FEMA
|
Back to Search Results |
|
Model Number M0068507000 |
Device Problem
Bent (1059)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/27/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a solyx sis system was used during sling procedure on (b)(6) 2016.According to the complainant, during the procedure, the tip of the device became bent.The procedure was completed with another solyx sis system.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|
Manufacturer Narrative
|
A visual examination of the returned solyx sis system revealed that the barbed delivery device tip is bent and the knob/tube assembly is pushed forward onto the bent tip and cannot be retracted.Analysis revealed that the mesh is stretched on one side.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.
|
|
Event Description
|
It was reported to boston scientific corporation that a solyx sis system was used during sling procedure on (b)(6) 2016.According to the complainant, during the procedure, the tip of the device became bent.The procedure was completed with another solyx sis system.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
|
|