• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK BIOTECH BIODESIGN POSTERIOR PELVIC FLOOR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK BIOTECH BIODESIGN POSTERIOR PELVIC FLOOR GRAFT Back to Search Results
Catalog Number J-PF-POST
Device Problem Insufficient Information (3190)
Patient Problems Emotional Changes (1831); Incontinence (1928); Unspecified Infection (1930); Pain (1994); Seroma (2069); Complaint, Ill-Defined (2331); Malaise (2359); Blood Loss (2597)
Event Type  Injury  
Manufacturer Narrative
Lot number lb56967c, as noted by the complainant, is not a valid cbi lot number.The reported lot number was unable to be narrowed down to a valid cbi lot number.This mdr is related to mdr 1835959-2016-00626.The lot number lb56967c, as noted by the complainant, is not a valid cbi lot number.The reported lot number was unable to be narrowed down to a valid cbi lot number.A review of the cbi complaint database did not reveal any previously reported complaints that match the details of this feedback.The ifu notes that "the following complications are possible with the use of surgical graft materials: bleeding, infection, adhesions, sterile effusion, chronic inflammation, allergic reaction, and delayed or failed incorporation of graft.If conditions of infection, inflammation, or allergic reaction cannot be resolved, consider removal of the graft." based on the information provided by the complainant, details regarding a specific correlation between the biodesign surgisis posterior pelvic floor graft's performance and the alleged injury remain unknown.A root cause of the allegations is inconclusive due to the lack of details provided by the complainant.
 
Event Description
The patient filed a device incident report to the (b)(6).This was documented as device incident report (b)(6).The patient reported she was implanted with two surgisis biodesign posterior pelvic floor grafts.One graft was placed in the anterior vaginal compartment and the other graft was placed in the posterior vaginal compartment.The date of implantation, surgical procedure performed, facility in which the surgery took place and the surgeon's name were not provided by the complainant.The patient reported experiencing continuous post-operative pain and a general feeling of being unwell.She also reported a discharge of blood and an offensive smell.The pain was described as sharp and stringing throughout the pelvic region.The patient also reported having persistent infections followed by skin changes, a loss of balance, and general malaise with the pain throughout the first twelve months after surgery.She indicated that the pain still persists, at the present time, and that the systems for which the procedure was performed to correct are worse than before surgery.The patient reports having permanent incontinence of both bowel and bladder and a foul smelling pelvic discharge (putrid seroma).She indicated she has had a loss of quality of life and is unable to have sexual intercourse.The patient reports that the products are unable to be removed and have caused irreparable internal damage.
 
Manufacturer Narrative
Date of event not provided by the complainant.Product common posterior pelvic floor graft; product code pag.Lot number lb56967c, as noted by the complainant, is not a valid cbi lot number.The reported lot number was unable to be narrowed down to a valid cbi lot number.Product expire date unknown; lot number not provided.Implant date reported as (b)(6) 2012.Product manufacture date unknown; lot number unknown.This mdr is related to mdr 1835959-2016-00626.The lot number lb56967c, as noted by the complainant, is not a valid cbi lot number.The reported lot number was unable to be narrowed down to a valid cbi lot number.A review of the cbi complaint database did not reveal any previously reported complaints that match the details of this feedback.The ifu notes that "the following complications are possible with the use of surgical graft materials: bleeding, infection, adhesions, sterile effusion, chronic inflammation, allergic reaction, and delayed or failed incorporation of graft.If conditions of infection, inflammation, or allergic reaction cannot be resolved, consider removal of the graft." based on the information provided by the complainant, details regarding a specific correlation between the biodesign surgisis posterior pelvic floor graft's performance and the alleged injury remain unknown.A root cause of the allegations is inconclusive due to the lack of details provided by the complainant.Update: the additional details received do not alter the previous investigation summary.
 
Event Description
Per device incident report (b)(4).Two x grafts were implanted in both anterior ((b)(4)) & posterior ((b)(4)) compartments of vagina.Pain in the vaginal region was continuous, general feeling unwell persisted continually after surgery.Having pain and continuously, discharge of blood tissue and offensive smell.Sharp, stringing pain throughout pelvic region.Persistent infections followed by skin changes and loss of balance, general malaise continuing with pain intensity throughout first 12 months continually - till present time.All of the systems which the operation was performed to correct are now worse than ever.Permanent incontinence from both bowel & urethra.Persistent pain associated with new formation of number of adhesions, product has integrated, with natural tissue causing all pelvic tissue to become necrotic causing continuous pelvis discharge with foul smell (putrid seroma).Loss of quality of life, unable to have sexual intercourse.Product is unable to be removed and has caused irreparable damage internally.Update: new information received which revealed the patient's identity, age at the time of biodesign surgisis posterior pelvic floor graft(s) implant, surgeon's name, geographical location implant took place, and date of implant.In (b)(6) 2012, the patient underwent surgery performed by dr.(b)(6) in (b)(6), for repair of minor incontinence and pelvic prolapse issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIODESIGN POSTERIOR PELVIC FLOOR GRAFT
Manufacturer (Section D)
COOK BIOTECH
1425 innovation place
west lafayette IN 47906
Manufacturer Contact
perry guinn
1425 innovation place
west lafayette, IN 47906
7654973355
MDR Report Key6053408
MDR Text Key58300324
Report Number1835959-2016-00627
Device Sequence Number1
Product Code PAG
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K992159
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberJ-PF-POST
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/07/2016
Initial Date FDA Received10/25/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/06/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
J-PF-POST
Patient Outcome(s) Disability;
Patient Age48 YR
-
-