• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER, INC. ZIMMER NATURAL NAIL CORTICAL SCREW; TRAUMA PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER, INC. ZIMMER NATURAL NAIL CORTICAL SCREW; TRAUMA PROSTHESIS Back to Search Results
Catalog Number 47248402750
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/26/2016
Event Type  malfunction  
Manufacturer Narrative
This report will be amended when our investigation is complete.
 
Event Description
It is reported the outer packaging was not sealed upon receipt.
 
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was not confirmed.Product was returned with the packaging in an opened condition; the tyvek on the outer cavity has been partially pulled open and is still attached to the thermal cavity, with evidence of proper sealing of the lid in the form of dry glue completely covering the exposed sealing edge.Device history record was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER NATURAL NAIL CORTICAL SCREW
Type of Device
TRAUMA PROSTHESIS
Manufacturer (Section D)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer (Section G)
ZIMMER, INC.
1800 west center street
warsaw IN 46580
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6053684
MDR Text Key58303379
Report Number0001822565-2016-03891
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK083497
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number47248402750
Device Lot Number63222817
Other Device ID NumberSEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2016
Initial Date FDA Received10/25/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-