Udi= (b)(4).Device evaluated by mfr.: returned product consisted of a ffr comet pressure wire in one piece.A visual inspection showed that the tip showed no issues.The shaft showed one kink approximately 128cm from the tip.The shaft showed some peeling of the ptfe coating approximately 129cm from the tip.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
|