Olympus followed up with the user facility multiple times to obtain additional information regarding the reported event but with no results.The customer¿s trocar was not returned to olympus for evaluation; however, the uhi-3 unit was returned.The reported event was not confirmed.A complete functionality inspection which included leak testing was performed on the customer¿s unit and no anomalies were found.Due to the customer¿s trocar not being returned the root cause of the reported event could not be determined.The instruction provides users several warnings in attempt to prevent patient complications.¿to avoid complications, monitor patient parameters such as endtidal co2, electrocardiogram, body temperature, etc., while using the uhi-3.Confirm that the suction tube is securely attached.If the suction tube becomes disconnected, co2 discharge will not be possible.In this case, further supply of co2 gas will increase the abdominal pressure.If a significant gas leak is noted from within the uhi-3, terminate its use immediately and contact olympus.¿.
|
Olympus was informed that during an unspecified procedure, a significant amount of co2 was injected into the patient (with bolus) resulting in an increased blood pressure.It was reported by the head of surgery that this was due to an air leaked on the insufflation unit's trocar.The intended procedure was completed and the patient's blood pressure was stabilized.Additionally, the insufflation unit was inspected post procedure by the user facility's biomed and he confirmed that the unit could not regulate the volume of gas due to the leak found on the trocar and that the unit compensated by sending out gas at a higher rate.
|