• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-920PA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-920PA Back to Search Results
Model Number ZM-920PA
Device Problems Overheating of Device (1437); Battery Problem (2885); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/30/2016
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the transmitter got very hot and was going through batteries quickly.The device was in use on a patient at the time, however no patient harm was reported.The customer will receive an exchange device.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the transmitter got very hot and was going through batteries quickly.
 
Manufacturer Narrative
Manufacturer narrative: the customer reported that the transmitter got very hot and was going through batteries quickly.The device was in use on a patient at the time, however no patient harm was reported.An exchange was sent to the customer.An evaluation was performed on the unit as well as an exterior visual inspection.Unit did not overheat when batteries were installed.Noted physical damage to battery compartment.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-920PA
Type of Device
ZM-920PA
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
486 nanokaichi
attn: shama mooman
gunma 370-2 343
JA   370-2343
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6062750
MDR Text Key58926205
Report Number8030229-2016-00535
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K945578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/27/2016,09/30/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-920PA
Device Catalogue NumberZM-920PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/27/2016
Distributor Facility Aware Date09/30/2016
Device Age130 MO
Event Location Hospital
Date Report to Manufacturer10/27/2016
Initial Date Manufacturer Received 10/27/2016
Initial Date FDA Received10/27/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-