• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Difficult to Interrogate (1331); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/29/2016
Event Type  malfunction  
Event Description
It was reported that usb cable connector to tablet was causing communication issue on programming system.It was reported that the tablet and the cable have to be on a level surface otherwise no interrogation is possible with the programming system.The suspect usb cable was received by the manufacturer.Analysis of the device is underway, but it has not been completed to date.Review of manufacturing records confirmed all tests passed for the tablet prior to distribution.
 
Event Description
An analysis was performed on the returned usb to db9 cable and the reported allegation was verified.The cause for the reported allegation is associated with an open wire connection in the returned serial cable.Once the wire was soldered onto the serial cable pcb, no further anomalies were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6063070
MDR Text Key58714091
Report Number1644487-2016-02496
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/05/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/05/2016
Initial Date FDA Received10/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-