Model Number 10094141
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/20/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a fd cooling unit failure due to improper workmanship.The affected couplings were exchanged and the system was brought back within specifications.Additionally, a corrective action was initiated to all affected systems and this action has been reported to the fda under 2240869-02/19/16-0006-c.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee biplane system.During a patient procedure, no x-ray was possible on plane a and the customer reported leaking liquid.The patient was safely removed from the system and transferred to an alternative system where the procedure was completed.We are unaware of any impact to the state of health of the patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE BIPLANE
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd
65-1a
malvern, PA 19355
6104486461
MDR Report Key6063188
MDR Text Key59022196
Report Number2240869-2016-05464
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K073290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial
Report Date 09/20/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10094141
Initial Date Manufacturer Received 09/20/2016
Initial Date FDA Received10/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-