Model Number 10280959
Device Problems Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2016
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.This event occurred in (b)(6).
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zeego system.During a patient procedure, no system movements were possible.The examination was aborted.We are unaware of any impact to the state of health of any patient involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.An investigation was conducted using the available log files.The root cause was determined to be a loose contact at the uli board (user location interface).The uli board was exchanged and the error did not reoccur.As this is the first known event of this nature, the event does not exceed the threshold in regards to the defect acceptance level and would not lead to a general field corrective action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEEGO
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6063200
MDR Text Key59041709
Report Number2240869-2016-05052
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K141574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 08/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number10280959
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/29/2016
Initial Date FDA Received10/28/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse