• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Device Problems Bent (1059); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/26/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # n92j4a.The analysis results found that the er320 device was received with no damage in the external components.In an attempt to replicate the reported incident, the device was tested for functionality; upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lock out after all the clips have been fired; therefore a potential cause of the customer reported experience is the firing of all of the clips and the instrument "will not fire" (activation of the lock out mechanism).No conclusion could be reached as to what may have caused the event reported.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during a laparoscopic cholecystectomy procedure, the clips jammed in the device.The device was removed from the trocar and the instrument jaws looked bent.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key6063980
MDR Text Key58766717
Report Number3005075853-2016-06155
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Catalogue NumberER320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/07/2016
Initial Date FDA Received10/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-