The patient was supported by an extracorporeal circulatory support device.It was reported that the patient was returned to the operating room (or) during the first postoperative night on (b)(6) 2016 for "clotted off lines." it was reported that the bedside nurse clamped the outlet and inlet blood tubing due to visualized clot with "layering/separating of blood" in the blood tubing.Flow of 0.0lpm was displayed on the device console.It was reported that the low flow alarm limit had not been set by hospital staff, and that the console was not alarming.The patient tolerated the disruption in support for a period of time; however, eventually became hemodynamically unstable.The patient was started on inotropic and vasoactive support due to mean arterial pressure between 20-30mmhg, and was returned to the operating room after exchanging the device and removing clot from the blood tubing, the patient stabilized.It was reported that there was not an issue with the extracorporeal circulatory support device or the primary console.The blood tubing had been caught under the patient's leg, and the tubing then clotted.The patient was brought back to the o.R.On (b)(6) 2016 to evacuate chest hematoma, and the patient was then weaned off of device support at that time.
|