SALIX PHARMACEUTICALS, INC. DEFLUX INJECTABLE GEL; AGENT,BULKING,INJECTABLE FOR GASTRO-UROLOGY USE Back to Search Results
Model Number DEFLUX
Device Problems Improper or Incorrect Procedure or Method (2017); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Renal Failure (2041); Obstruction/Occlusion (2422)
Event Type  Injury  
Manufacturer Narrative
Additional information: investigation of this event is in progress.A follow-up report will be submitted upon completion of the investigation.
 
Event Description
Literature article "delayed obstruction with asymptomatic loss of renal function after dextranomer/hyaluronic acid copolymer (deflux) injection for vesicoureteral reflux: a close look at a disturbing outcome" by dimitri papagiannopoulos, ilina rosoklija, earl cheng, elizabeth yerkes, reports: "three cases where deflux was used in off-label fashion, resulting in delayed ureteral obstruction and loss of renal function (range 18-52 months post-operatively).We now place increased emphasis on the need for long-term follow-up after deflux in both routine and complex cases, particularly in situations of off-label use." case one of three: a (b)(6) patient with neurogenic bladder, a notable decline in renal function and renal parenchymal loss on the left kidney (vcug grade: 3, pelviectasis and dmsa: 27%df) was treated with double hit deflux injection.During intraoperative observations the surgeon noticed distally ectopic but intravesical ureter.Patient developed febrile uti and chronic urothelial inflammation after 18 months.Ultra sonography showed left kidney grade 4 hun, and dmsa: 16% df.Patient was treated with open ureteral re-implantation with extravesical dissection.
 
Manufacturer Narrative
During an internal review, it was identified that the medwatch report for this event was electronically submitted to fda with an mdr number that was generated using (b)(4) as the registration number.The correct registration number that should have been used to generate the mdr number is (b)(4).
 
Manufacturer Narrative
The device is not available to return for evaluation, and the lot number was not provided; therefore a dhr review could not be performed.Based on the current information the root cause of the event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEFLUX INJECTABLE GEL
Type of Device
AGENT,BULKING,INJECTABLE FOR GASTRO-UROLOGY USE
Manufacturer (Section D)
SALIX PHARMACEUTICALS, INC.
raleigh NC
Manufacturer (Section G)
Q-MED
seminariegatan 21
uppsala 75228
SW   75228
Manufacturer Contact
faranak gomarooni
50 technology drive west
irvine, CA 92618
9493985708
MDR Report Key6067922
MDR Text Key58795967
Report Number0001313525-2016-00500
Device Sequence Number1
Product Code LNM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type literature
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/04/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberDEFLUX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/04/2016
Initial Date FDA Received10/31/2016
Supplement Dates Manufacturer ReceivedNot provided
10/04/2016
Supplement Dates FDA Received02/14/2017
10/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age3 YR
-
-