• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.00021
Device Problems Overheating of Device (1437); Battery Problem (2885)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 10/03/2016
Event Type  malfunction  
Manufacturer Narrative
The device has been returned and an evaluation is pending.A follow-up report will be submitted upon evaluation completion.Aribex has issued a recall, reference (b)(4).
 
Event Description
It was reported that battery is getting hot when placed on the charging cradle.
 
Manufacturer Narrative
Upon visual inspection, it appears that there was a failure to the control board and not a thermal runaway event.This concludes the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO 2
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6069792
MDR Text Key59291115
Report Number1017522-2016-00036
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 11/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.00021
Device Catalogue Number0.850.00021
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/10/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/03/2016
Initial Date FDA Received11/01/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-