• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. ATTUNE TIBIAL TRIAL EXTRACTOR; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. ATTUNE TIBIAL TRIAL EXTRACTOR; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 254500138
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 10/25/2016
Event Type  malfunction  
Manufacturer Narrative
This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The instrument broke while surgeon was trying to extract the poly trial.One of the prongs broke off the extractor.
 
Manufacturer Narrative
Examination of the returned device confirms the reported event of breakage.(b)(4) was held on 25 jan 2016 to evaluate the increased complaints involving the attune tibial trial extactor (product code 254500138).The review team determined that there was no increased patient risk.Based on a previous evaluation ((b)(4)) for a similar failure by depuy material science, the root cause is attributed to overload.Based on the root cause of overload and (b)(4) determination of no additional patient risk, corrective action was not indicated.Continue to monitor via (b)(4).Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE TIBIAL TRIAL EXTRACTOR
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6072078
MDR Text Key58929212
Report Number1818910-2016-30630
Device Sequence Number1
Product Code HWB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500138
Device Lot NumberBFA0NRV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/25/2016
Initial Date FDA Received11/02/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/23/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
-
-