The manufacturing batch lot was not provided therefore we were unable to review the device history records to confirm that the device met all material, assembly and inspection specifications prior to shipment.The device was received for analysis on 11/17/2016.As received, the specimen consists of one each clinically used, damaged, unknown safari wire; returned coiled, loose and double bagged within "zip-lock" style poly biohazard pouches.Due to the fact that the catheter was not returned with the guidewire the complaint could not be confirmed.The returned specimen presents several kinks/bends and several regions of offset coil damage scattered throughout the length of the device.Frayed/scraped ptfe coating is present adjacent to the kink/bend damage.The offset coil damage creates localized oversized diameters to.03865".The specimen also presents several stretched coil regions with the coil displacement in a proximal to distal orientation.Dimensionally, the specimen appears to be a best fit match to a safari2 275cm exsml crv wire.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the evidence presented by the sample and the information provided by the supporting documentation, clinical and/or procedural factors appear to have impacted on the event as reported.Based on the returned guidewire being identified as a possible best fit for a safari2 the 510k number was updated in this medwatch report.
|