The customer was contacted on (b)(6) 2016 and it was determined that the customer had an issue with patient sample results due to an issue with assay cups used on the e602 analyzer.The assay cups had holes in them.The customer stated alarms occurred on the analyzer on (b)(6) 2016.A clinician contacted the laboratory on (b)(6) 2016 explaining a patient result was not possible.Based on the alarms and the possibility the cups affected the results, the customer decided to repeat samples that were initially tested around the time of the alarms.Of the repeated samples, 5 had questionable results.Of the 5 samples, 3 had erroneous results that were reported outside of the laboratory for the elecsys tsh assay (tsh).All 3 samples were repeated on (b)(6) 2016 after changing the assay cups.The customer called the clinician to amend the results.The first sample initially resulted as <0.005 ui/ml.The sample was repeated, resulting as 1.11 ui/ml.The second sample initially resulted as <0.005 ui/ml.The sample was repeated, resulting as 1.75 ui/ml.The third sample initially resulted as <0.005 ui/ml on (b)(6) 2016.The sample was repeated, resulting as 1.33 ui/ml.The patients were not adversely affected.The tsh reagent lot number and expiration date were requested but not provided.The e602 analyzer serial number was requested but not provided.The laboratory is no longer using the lot in question.
|
The field service engineer cleaned the analyzer and the laboratory removed all assay cups of lot 16418713.A specific root cause could not be determined based on the provided information.Additional information required for the investigation was requested, but not provided.An issue with sample handling or an instrument issue could not be excluded.It is very likely that the observed issue was caused by damaged assay cups of lot 16418713.Damaged assay cups of lot 16418713 have been confirmed.As a workaround, other lot numbers are available for use.
|