Model Number CTRF100 |
Device Problem
Energy Output Problem (1431)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/07/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported condition was confirmed.The investigation isolated the failure to the audio board, but a root cause was not identified.The probable root cause could not be determined based on the information provided.The audio board was replaced to address the condition.No trend has been identified.
|
|
Event Description
|
The customer reported that during an rf ablation procedure for liver cancer of 5mm s8 tumor, the unit was powered on and the needle was punctured.Resistance value was shown at 100 ohm but the output was below 7w and an error occurred.The needle was replaced but the incident occurred again.The patient was scheduled for a hepatectomy via open procedure.The team was going to ablate during the open procedure when the generator failed.
|
|
Manufacturer Narrative
|
One cool-tip audio board was received for evaluation.A visual inspection of the audio board found the torque seal on potentiometers p1, p2, p3, p4, and p7 through p12 had been removed.This may indicate that the potentiometers factory settings had been altered.The visual inspection found no other anomalies.The audio board was placed in cool-tip test unit.At initial power up, the test unit failed self-test by producing an impedance test failure.An impedance calibration was performed.At next power up the test unit passed self-testing.Testing found that the board would not accept calibration.Troubleshooting could not determine the cause of the calibration failures.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The customer reported that during an rf ablation procedure for liver cancer of 5mm s8 tumor, the unit was powered on and the needle was punctured.Resistance value was shown at 100 ohm but the output was below 7w and an error occurred.The needle was replaced but the incident occurred again.The patient was scheduled for a hepatectomy via open procedure.The team was going to ablate during the open procedure when the generator failed.
|
|
Search Alerts/Recalls
|