• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problems Computer Software Problem (1112); Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/24/2014
Event Type  malfunction  
Event Description
During an internal review of programming and diagnostic history, it was identified that an interrupted system diagnostic test occurred that caused an unintended change in device settings during an office visit on (b)(6) 2014.The settings were not corrected on the same date.The device settings were ok on (b)(6) 2015.The review of the manufacturing records confirmed that the programming computer passed all functional tests prior to the distribution.No patient adverse events were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6078543
MDR Text Key59520043
Report Number1644487-2016-02444
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063825
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 10/17/2016
Initial Date FDA Received11/04/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age20 YR
-
-