• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problem Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/10/2016
Event Type  malfunction  
Manufacturer Narrative
 
Event Description
It was reported that a physician's tablet could not be used to interrogate the device.Trouble-shooting confirmed that the programming system's wand and serial cable were working.Replacing the tablet resolved the problem.The issue was thought to be a loose connection at the tablet's usb port.Trauma to the tablet was denied.The suspect device has not been received by the manufacturing facility to date.No additional relevant information has been received to date.
 
Event Description
The suspect device was received and underwent product analysis.Product analysis confirmed that the usb port was damaged.The damage was due to mechanical stress.No other anomalies were identified.No additional relevant information has been received, to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6078587
MDR Text Key59510995
Report Number1644487-2016-02535
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Other Device ID NumberVERSION 11.0.5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2016
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 10/10/2016
Initial Date FDA Received11/04/2016
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/02/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-