This report or other information submitted by joerns healthcare under 21 cfr part 803, and any release by fda of that report information, does not reflect a conclusion or admission by joerns healthcare, its employees, its contract service firms, or their employees, finished device suppliers, or their employees caused or contributed to the reportable event.
|
It was reported to the manufacturer by the end user, per the end user, wound nurse went into the room and the mattress was deflated.The patient was lying on his side.The patient has a trachea and cannot move by himself.The mattress was slightly covering the trachea so the patient could not take full breaths.Additionally, when the mattress deflated, during numerous occasions, this caused bedsores on the patients' tailbone.The facility has tapped the hose assembly back together and is continuing to use under the resident.The bedsores were treated at the facility.Complaint#(b)(4) were entered into our system to have the mattress and control unit returned to joerns for investigation.As of this writing, the mattress and control unit have not been returned.
|