The cause of the discrepant falsely depressed d-dimer result is unknown.However, analysis of the instrument data revealed no systematic failures or qc failures.The issue is an isolated observation of a patient measurement with a falsely low d-dimer result in a single determination which is discordant to other measurements.The cause of the issue seems to be related to the affected 2nd sample.This sample is described as having a high count of fibrin monomers.Fibrin monomers are an intermediates in the fibrinogen activation to fibrin and may cause microclots.Samples with microclots may cause sample under-aspiration by pipettor clogging and by this may deliver false low results.Microclots can be removed by correct sample preparation with centrifugation, and therefore the presence of microclots is a preanalytical issue which is under the user's responsibility.In addition, as the patient has a known malignancy suspected for dic/sepsis, the results of such samples have to be interpreted with caution as these disease stages are known to potentially interfere with d-dimer analysis as indicated in the 'summary and explanation' section of the instructions for use.The device is performing within specifications.No further evaluation of the device is required.
|