Model Number MMT-523NAP
Device Problems Device Displays Incorrect Message (2591); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/12/2016
Event Type  malfunction  
Manufacturer Narrative
The insulin pump had compromised force sensor system alarm during basic occlusion test due to cracked end cap.Analysis was unable to complete functional test due to compromised force sensor system alarm.The insulin pump was received with minor scratched lcd window, broken reservoir tube lip and cracked battery tube threads.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed (b)(4).
 
Event Description
The customer reported via phone call that they received compromised force sensor system alarm.The customer's blood glucose was 120 mg/dl at the time of incident.The customer stated that the end cap was coming apart from the insulin pump.The customer was advised that the insulin pump will need to be replaced.The customer was advised to disconnect from the insulin pump and revert to back-up plan.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire street
northridge CA 91325 1219
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6091195
MDR Text Key59854906
Report Number2032227-2016-40962
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523NAP
Device Catalogue NumberMMT-523NAP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/17/2016
Initial Date FDA Received11/09/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age56 YR
Patient Weight68
-
-