(b)(6).Part 04.641.008, lot: 6963216: manufacturing location: (b)(4).Release to warehouse: june 20, 2012.No non-conformance reports were generated during production that would contribute to the complaint condition.The review of the device history record shows that there were no issues during the manufacture of the product that would contribute to the complaint condition.A product investigation was completed: one (1) small titanium rib hook (part 04.641.002, lot 6123429, manufacture date april 13, 2009) and one (1) small ti rib hook cap (part 04.641.008, lot 6963216, manufacture date june 20, 2012) were returned with a complaint stating the rib hook migrated from the rib postoperatively.The devices were returned joined together by a distraction lock (497.125, lot 9832340).The devices exhibited scratches and discoloration indicative of post-manufacturing damage due to implantation/explantation.Since the devices were returned assembled without any significant damage outside of normal use, it is unlikely that the migration was as a direct result of a malfunction of the devices.It is also uncertain how long the device was implanted.The complaint of migration could not be confirmed with the given information.Although a definitive root cause could not be determined, stresses imposed on the device due to patient activity, patient growth, and/or improper tightening during implantation.A visual inspection, complaint history review, drawing review, and risk assessment review were performed as part of this investigation.Four (4) ti distraction locks (part 497.125; lots 9832340, 6409102, 7535699, 7004950) were returned as concomitant devices without an alleged complaint condition.Upon visual inspection, there is no evidence that these devices contributed to the complaint condition.Additionally, lot 9832340 that was returned assembled to the complaint devices is considered concomitant as its function is to hold the rib hook and the cap together and would not directly contribute to the complaint condition of malfunction.The relevant product drawings were reviewed during the investigation.No drawing issues or discrepancies were noted.The design is adequate for its intended use and did not contribute to this complaint condition.No non-conformance reports were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Device report from synthes canada reported that a patient was originally implanted with vertical expandable prosthetic titanium rib ii (veptr ii) on an unknown date.Postoperative x-ray taken on an unknown date revealed that rib hook had been migrated from the rib.On (b)(6) 2016 the vertical expandable prosthetic titanium rib ii (veptr ii) revision surgery performed as the part of the planned treatment and also due to previous rib hook had migrated from the rib.Patient was implanted with four (4) new rods, three (3) new rib hooks, three (3) new rib hook caps and five (5) new distraction locking caps.Surgery was completed successfully with no other medical intervention required.Patient status reported as stable after the surgery.Concomitant medical products: veptr rods distal and proximal (part unknown, lot unknown, quantity 4); rib hooks (part unknown, lot unknown, quantity 2); rib hook caps (part unknown, lot unknown, quantity 2); distraction locking caps (part 497.125, lot 6409102, 7535699, 9832340, quantity 3).There were several intra-operative issues which caused 45 minutes delay during the procedure; these are captured in linked complaint (b)(4).This is report 2 of 2 for (b)(4).
|