Model Number WCD 4000
Device Problem Cut In Material (2454)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/07/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of monitor sn (b)(4) has been completed.The reported problem (exposed wires) was confirmed.Upon investigation the monitor was unable to communicate with an electrode belt.The monitor's electrode belt connector was broken free from the monitor enclosure, damaging wires within the connector.The root cause for the damaged connector could not be positively identified.No adverse event resulted from the defective monitor.
 
Event Description
A us distributor returned a patient's monitor and reported that the monitor had exposed wires.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
zachary nelson
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key6094149
MDR Text Key60026331
Report Number3008642652-2016-08243
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 11/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2016
Initial Date Manufacturer Received 11/01/2016
Initial Date FDA Received11/10/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-