The medtronic representative confirmed that the same visualase thermal therapy system and scanner were used in another procedure performed hours before, in the same room, with no issues.On 10/25/2016 medtronic representative reported that the background is regarding a large body habitus patient that had an occipital approach.The primary scanner coil would not afford the patient to fit without damaging the fiber.We switched to that rooms only other option, which was a 4-channel transmit only body coil.Although the tmap represented changes in artifact temperature variations, it refused to display temperature variations surrounding the fiber.After trouble-shooting steps and process of elimination, we transferred the patient to their secondary scanner, which has a better head coil that is clamshell and separates into two halves.On 11/08/2016 a medtronic representative, following-up at the site, reported the image was normal in appearance and some thermography noise could be seen, just no change in temperature.It was definitely a very old body coil that was used.On 10/31/2016 a medtronic representative performed a thermal therapy system check-out, all areas passed.System performed as intended.
|
A medtronic representative reported that, while in a laser induced thermal therapy (litt) procedure, they were unable to see any temperature change on the t-map.This occurred while only at a test dose wattage.There was no heat ablation shown on the monitor screen.In trouble-shooting, confirmed the indication light showed that the disposable was connected.They were using a body coil, transceiver only.Confirmed good connection to the mri machine (ge scanner).The medtronic representative re-started the session and re-started software without resolution.The surgeon opted to continue and completed the procedure with the use of the thermal therapy system.Delay in therapy was greater than one hour before continuing.There was no impact on patient outcome.
|